SYSTem Subsystem
System commands control system functions that are not directly related to output control,
measurement, or status functions. Note that IEEE-488 Common commands also control system
functions such as self-test, saving and recalling states, and others.
SYSTem:COMMunicate:LAN:CONTrol?
SYSTem:COMMunicate:TCPip:CONTrol?
Returns the initial socket control connection port number. This connection is used to send and receive
commands and queries. Unlike the data socket, which uses a fixed port number, the control socket
port number varies and must be obtained using these queries.
Parameter
Typical Return
(none)
<port #>
(0 if sockets not supported)
Query the Control connection port number:
SYST:COMM:LAN:CONT? or SYST:COMM:TCP:CONT?
SYSTem:COMMunicate:RLSTate LOCal|REMote|RWLock
SYSTem:COMMunicate:RLSTate?
Configures the remote/local state of the instrument. Remote and local do the same thing and are
included for compatibility with other products. LOCal sets instrument to front panel control. REMote
sets the instrument to front panel control. RWLock disables the front panel keys. The instrument can
only be controlled via the remote interface. This programmable setting is completely independent
from the front panel lock/unlock function.
Parameter
Typical Return
LOCal|REMote|RWLock , *RST LOCal
LOC, REM, or RWL
Sets the remote/local state to remote: SYST:COMM:RLST REM
l
The remote/local state is unaffected by *RST or any SCPI commands other than SYSTem:COMMu-
nicate:RLState.
l
The remote/local instrument state can also be set by other interface commands over the GPIB
and some other I/O interface.
l
When multiple remote programming interfaces are active, the interface with the most recently
changed remote/local state determines the instrument’s remote/local state.
Keysight N69100 Series Operating and Service Guide
197
5 SCPI Programming Reference
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......