Parameter
Typical Return
DIO | DINPut | FAULt | INHibit | ONCouple |
OFFCouple TINPut | TOUTput
DIO, DINP, FAUL, INH, ONC, OFFC, TINP,
or TOUT
Sets pin 1 to FAULt mode: DIG:PIN1:FUNC FAUL
[SOURce:]DIGital:PIN<1-7>:POLarity POSitive|NEGative
[SOURce:]DIGital:PIN<1-7>:POLarity?
Sets the pin polarity.
POSitive means a logical true signal is a voltage high at the pin. For trigger inputs and outputs,
POSitive means a rising edge.
NEGative means a logical true signal is a voltage low at the pin. For trigger inputs and outputs,
NEGative means a falling edge. The pin polarities are saved in non-volatile memory.
Parameter
Typical Return
POSitive|NEGative
POS or NEG
Sets pin 1 to POSitive polarity: DIG:PIN1:POL POS
[SOURce:]DIGital:TOUTput:BUS[:ENABle] 0|OFF|1|ON
[SOURce:]DIGital:TOUTput:BUS[:ENABle]?
Enables or disables BUS triggers on digital port pins. This allows a BUS trigger to be sent to any digital
port pin that has been configured as a trigger output. A trigger out pulse is generated when the state is
on and a bus trigger is received. A BUS trigger is generated using the *TRG command.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Enable BUS triggered signals on the digital pins: DIG:TOUT:BUS ON
l
The query returns 0 (OFF) if the trigger signal will NOT be generated with a BUS trigger command,
and 1(ON) if a trigger signal will be generated with a BUS trigger command.
Keysight N69100 Series Operating and Service Guide
163
5 SCPI Programming Reference
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......