CALibrate:STATe 0|OFF|1|ON [,<password>]
CALibrate:STATe?
Enables or disables calibration mode. Calibration mode must be enabled for the instrument to accept
any calibration commands. The first parameter specifies the state. The second optional parameter is
the password.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
<
password
> a numeric value up to 15 digits
(none)
Disable calibration: CAL:STAT OFF
Enable calibration: CAL:STAT ON [,value]
A password is required if <
password
> has been set to a non-zero value.
CALibrate:VOLTage[:LEVel] <
value
>
Calibrates the local voltage programming and measurement. The value selects the range to
calibrate.
Parameter
Typical Return
The maximum voltage of the output range.
(none)
Calibrates the voltage of the 20 V range: CAL:VOLT 20
5 SCPI Programming Reference
158
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......