![Keysight Technologies N69100 Series Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218024.webp)
Command Quick Reference
Some [optional] commands have been included for clarity. All settings commands have a
corresponding query. See the
for SCPI.
:ACQuire
Cancels any triggered measurements.
:TRANsient
Cancels any transients or Arbs.
:COUNt?
Returns the number of times the unit has been calibrated.
:CURRent
[:LEVel] <
value
>
Calibrates the current programming and measurement.
:DATA <
value
>
Enters the calibration value read by the external meter.
:DATE <
"date"
>
Enters the calibration date in nonvolatile memory.
:LEVel P1|P2|P3
Advances to the next level in the calibration.
:PASSword <
value
>
Sets a numeric password to prevent unauthorized calibration.
:SAVE
Saves the calibration constants in non-volatile memory.
:STATe 0|OFF|1|ON
Enables or disables calibration mode.
:VOLTage
[:LEVel] <
value
>
Calibrates the local voltage programming and measurement.
[:WINDow]
[:STATe] 0|OFF|1|ON
Turns the front panel display on or off.
:VIEW METER_VI|METER_VP|METER_VIP
Selects the parameters to display on the front panel.
:SAVer
[:STATe] 0|OFF|1|ON
Turns the front panel screen saver on or off.
[:SCALar]
:CURRent
[:DC]?
Returns the averaged measurement.
:ACDC?
Returns the RMS measurement (AC + DC).
:HIGH?
Returns the High level of a pulse waveform.
:LOW?
Returns the Low level of a pulse waveform.
:MAXimum?
Returns the maximum value
:MINimum?
Returns the minimum value
:POWer
[:DC]?
Returns the averaged measurement.
:MAXimum?
Returns the maximum value
:MINimum?
Returns the minimum value
:VOLTage
[:DC]?
Returns the averaged measurement.
:ACDC?
Returns the RMS measurement (AC + DC).
1 Quick Reference
24
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......