![Keysight Technologies N69100 Series Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218034.webp)
Supplemental Characteristics
Supplemental characteristics are not warranted but are descriptions of performance determined
either by design or by type testing. Supplemental characteristics are typical unless otherwise noted.
Characteristic
N69143A
N69145A
N69146A
N69172A
N69173A
Voltage programming
Range:
Resolution:
0.02 to 20.4 V
3.06 mV
0.08 to 81.6 V
12.8 mV
0.16 to 163.2V
25.6 mV
1 to 1020 V
160 mV
2 to 2040 V
320 mV
Current programming
Range:
Resolution:
0 to 816 A
248 mA
0 to 255 A
80 mA
0 to 127.5A
40 mA
0 to 61.2A
19.2 mA
0 to 30.6A
9.6 mA
Resistance programming
Range:
Resolution:
Accuracy:
0 to 0.049
Ω
0.4 µ
Ω
0.05% + 2 µ
Ω
0 to 0.625
Ω
4.8 µ
Ω
0.05% + 16 µ
Ω
0 to 2.5
Ω
19.6 µ
Ω
0.05% + 50 µ
Ω
0 to 69 Ω
0.55 mΩ
0.06% + 0.55 mΩ
0 to 277 Ω
2.2 mΩ
0.06% + 2.2 mΩ
Additional V-programming offset/
Ω
:
100 mV
30 mV
15 mV
N/A
N/A
Measurement range
Voltage:
Current:
0 to 25 V
-202.8 A to 2028 A
0 to 100 V
-63 A to 630 A
0 to 200 V
-31.5 A to 315 A
0 to 1200 V
-15.3 A to 153 A
0 to 2400 V
-7.65 A to 76.5 A
Program & measure tempCo/°C 1
Voltage:
Current:
0.0015% +80 µV
0.0075% +2.3mA
0.0015% +300µV
0.0025% +800µA
0.0015%+600µV
0.0025%+400µA
0.002%+7 mV
0.005%+240µA
0.002%+14 mV
0.005%+120µA
Output ripple & noise (20 Hz to 20 kHz)
CC rms:
200 mA
70 mA
50 mA
30 mA
15 mA
Common mode current
rms:
peak-to-peak:
2 mA
20 mA
2.5 mA
24 mA
3.5 mA
30 mA
2.5 mA
30 mA
2.5 mA
30 mA
Overvoltage protection
Maximum setting:
Accuracy:
Response time:2
24 V
0.02% + 2 mV
< 30 µs
96 V
0.02% + 8 mV
< 30 µs
192 V
0.02% + 16 mV
< 30 µs
1200 V
0.02% + 200 mV
< 30 µs
2400 V
0.02% + 200 mV
< 30 µs
Voltage up/down programming, comp 0 3
Rise/fall time 10% to 90% of step:
Settling time to 0.1% of step:
10 ms
50 ms
10 ms
50 ms
10 ms
50 ms
10 ms
50 ms
10 ms
50 ms
Voltage up/down programming, comp 1 3
Rise/fall time 10% to 90% of step:
Settling time to 0.1% of step:
20 ms
100 ms
20 ms
100 ms
20 ms
100 ms
20 ms
100 ms
20 ms
100 ms
Voltage up/down programming, comp 2 3
Rise/fall time 10% to 90% of step:
Settling time to 0.1% of step:
40 ms
200 ms
40 ms
200 ms
40 ms
200 ms
40 ms
200 ms
40 ms
200 ms
Current up/down programming, comp 0 4
Rise/fall time 10% to 90% of step:
Settling time to 1% of step:
2.5 ms
5 ms
2.5 ms
5 ms
2.5 ms
5 ms
2.5 ms
5 ms
2.5 ms
5 ms
Current up/down programming, comp 1 4
Rise/fall time 10% to 90% of step:
Settling time to 1% of step:
5 ms
10 ms
5 ms
10 ms
5 ms
10 ms
N/A
N/A
1 Quick Reference
34
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......