Set the Output Voltage
When the unit is in voltage priority mode, the output voltage remains at its programmed setting as
long as the load current remains within its programmed positive or negative limit.
Front Panel Menu Reference
SCPI Command
Press the
Voltage
key.
Enter a value and press Select.
To set the output voltage to 400 volts:
When the unit is in current priority mode, you can specify a voltage limit which limits the output voltage
at the specified value. The output current remains at its programmed setting, provided the load
voltage remains within the voltage limit setting.
Front Panel Menu Reference
SCPI Command
Press the
Voltage
key.
Specify a + Voltage limit. Then press
Select
.
To set the voltage limit:
Set the Output Current
When the unit is in voltage priority mode, you can specify a positive and negative current limit, which
limits the output current at the specified value.
Front Panel Menu Reference
SCPI Command
Press the
Current
key.
Specify a positive or negative Current limit.
Then press
Select
.
To set the positive current limit:
To set the negative current limit:
When the unit is in current priority mode, you can specify a positive or negative output current level,
which will be maintained as long as the output voltage remains within its programmed limit.
Front Panel Menu Reference
SCPI Command
Press the
Current
key.
Enter a positive or negative value.
Then press
Select
.
To set the current to +5 amperes:
To set the current to –5 amperes:
Set the Slew Rate
The voltage slew rate determines the rate at which the voltage changes to a new setting. This only
applies to voltage settings while operating in voltage priority mode. When set to MAXimum, INFinity,
or to a very large value, the slew rate will be limited by the unit's listed programming speed and
bandwidth. This setting can be used to prevent crossover into current limit while up- and down-
programming capacitive loads, or to limit programming response to a controlled rate. Use the
Keysight N69100 Series Operating and Service Guide
97
4 Using the DC Power System
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......