Instrument Sanitize
This procedure is not recommended for use in routine applications because of the
possibility of unintended loss of data.
This procedure sanitizes all user data. It writes all zeros to flash memory and then performs a full chip
erase as per the manufacturer’s data sheet. Identification data such as instrument firmware, model
number, serial number, MAC address, and calibration data is not erased. After the data is cleared, the
instrument is rebooted.
If you cannot access the Admin menu, it may be password protected.
Front Panel Menu Reference
SCPI Command
Select
System\Admin\Sanitize
Select Sanitize.
Selecting Sanitize removes all user-data from the
instrument and cycles power.
Refer to
http://rfmw.em.keysight.com/aerospace/index.aspx
for detailed information.
Turn-on after Sanitization
The first time the unit is turned on after it has been sanitized, several NVRAM checksum errors will be
generated. These errors annunciate the fact that two files were missing, which have been recreated
with default values. Reboot the unit one more time to remove the errors.
Sanitizing an Inoperative Instrument
If the instrument is no longer functioning and you are unable to use the sanitization procedure
described above, you must physically remove the P600 board from the instrument and destroy it.
Refer to
7 Service and Maintenance
250
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......