![Keysight Technologies N69100 Series Скачать руководство пользователя страница 240](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218240.webp)
Keysight N69145A
N69145A
Test Record
Report Number ______
Date__________
Test Description
Model
Min. Specs
Results
Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
0.066984
Vout - 0.013016
79.971
Vout - 0.029
_______
_______
_______
_______
0.093016
Vout + 0.013016
80.029
Vout + 0.029
CV Load Effect:
Both
- 0.003
_______
+ 0.003
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
0.080
0.008
Transient response @ 900 µs:
Both
- 0.8
_______
+ 0.8
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
- 0.115
Iout - 0.115
249.735
Iout - 0.265
_______
_______
_______
_______
0.115
Iout + 0.115
250.265
Iout + 0.265
CC Load Effect:
Both
- 0.025
_______
+ 0.025
Current Sink Tests
10% of current rating:
Both
- 25.6303
_______
- 25.3697
N69145A UUT Settings
RP7945A Load Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
10% of current rating (Isink):
Voltage priority: 0.08 V, 127.5 A
Voltage priority: 80 V, 127.5 A
Voltage priority: 40 V, 255 A
Voltage priority: 40 V, 255 A
Current priority: 40.4 V, 0 A
Current priority: 40.4 V, 250 A
Current priority: 80.8 V, 125 A
Voltage priority: 0 V, 0 A
not used
not used
Current priority: 40.4 V, -250 A
Current priority: 60 V, -100 A to -225 A, slew=4 A/µs
Voltage priority: 40 V, 0 A
Voltage priority: 40 V, -255 A
Voltage priority: 80 V, -127.5 A
Voltage priority: 40 V, 250 A
6 Verification and Calibration
240
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......