[SOURce:]CURRent:MODE FIXed|STEP|LIST|ARB
[SOURce:]CURRent:MODE?
Sets the transient mode. This determines what happens to the output current when the transient
system is initiated and triggered.
FIXed keeps the output current at its immediate value.
STEP steps the output to the triggered level when a trigger occurs.
LIST causes the output to follow the list values when a trigger occurs.
ARB causes the output to follow the arbitrary waveform values when a trigger occurs.
Parameter
Typical Return
FIXed|STEP|LIST|ARB, *RST FIXed
FIX, STEP, LIST, or ARB
Sets the current mode to Step: CURR:MODE STEP
[SOURce:]CURRent:PROTection:DELay[:TIME] <
value
>|MIN|MAX
[SOURce:]CURRent:PROTection:DELay[:TIME]? [MIN|MAX]
Sets the over-current protection delay. The over-current protection function will not be triggered
during the delay time. After the delay time has expired, the over-current protection function will be
active. This prevents momentary changes in output status from triggering the over-current protection
function. Values up to 255 milliseconds can be programmed, with a resolution of 1 millisecond.
Parameter
Typical Return
0 – 0.255, *RST 0.020 s
<delay value>
Sets the protection delay to 0.2 seconds: CURR:PROT:DEL 0.2
l
The operation of over-current protection is affected by the setting of the current protection delay
start event, which is specified by CURRent:PROTection:DELay:STARt.
[SOURce:]CURRent:PROTection:DELay:STARt SCHange|CCTRans
[SOURce:]CURRent:PROTection:DELay:STARt?
Specifies what starts the over-current protection delay timer.
SCHange starts the over-current delay whenever a command changes the output voltage, current, or
output state.
CCTRans starts the over-current delay timer by any transition of the output into current limit mode.
Parameter
Typical Return
SCHange|CCTRans, *RST SCHange
SCH or CCTR
Selects the CCTRans delay mode: CURR:PROT:DEL:STAR CCTR
5 SCPI Programming Reference
160
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......