![Keysight Technologies N69100 Series Скачать руководство пользователя страница 186](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218186.webp)
Parameter
Typical Return
0 - 1023, *RST 0
<delay value>
Sets a delay of 0.5 s before turning the output on: OUTP:DEL:RISE 0.5
l
Each instrument exhibits a minimum delay offset that applies from the time that a command to
turn on the output is received until the output actually turns on. If you specify a turn-on delay, this
delay will be added to the minimum delay offset, resulting in a turn-on delay that is actually longer
than the one you programmed.
l
Use OUTput:COUPle:MAX:DOFFset? to query the delay offset that is required for each instrument.
OUTPut[:STATe]:TMODE:COUPle 0|OFF|1|ON
OUTPut:[:STATe]TMODE:COUPle?
When coupling is enabled, changing the turn-on setting also changes the turn-off setting and vice-
versa. If the turn-on and turn-off settings are not the same and coupling is enabled, the turn-off
setting will be changed to agree with the turn-on setting.
Parameter
Typical Return
0|OFF|1|ON
0 or 1
Couples the Turn-on/turn/off modes: OUTP:REL:LOCK ON
OUTPut[:STATe]:TMODe[:OFF] HIGHZ | LOWZ
OUTPut:TMODe[:OFF]?
OUTPut[:STATe]:TMODe:ON HIGHZ | LOWZ
OUTPut:TMODe:ON?
These commands set the output turn-on and turn-off behavior to Low impedance or High impedance.
Low Impedance - programs the output voltage to zero, then disconnects the output. Maximum
negative current sinking occurs for up to 250 ms during the turn-off transition.
High Impedance - disconnects the output without actively sinking current.
Parameter
Typical Return
HIGHZ|LOWZ
HIGHZ or LOWZ
Sets the turn-on state to high impedance: OUTP:TMOD:ON HIGHZ
Sets the turn-off state to low impedance: OUTP:TMOD:OFF LOWZ
l
The turn-on/turn-off setting only applies when the instrument is operating in voltage priority
mode. In current priority mode, the turn-on/turn-off behavior is always high impedance.
5 SCPI Programming Reference
186
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......