Test Considerations
For optimum performance, all verification procedures should comply with the following:
l
Ambient temperature is stable, within 23 °C ±2°C.
l
Ambient relative humidity is less than 80%.
l
30 minute warm-up period before verification or adjustment.
l
Cables as short as possible, twisted or shielded to reduce noise.
Verification Procedure
Voltage Programming and Readback Accuracy
This test verifies that the voltage programming and measurement functions are within specifications.
Step 1. Turn off the unit under test. Only connect the DMM to the sense screws (see
). Do
NOT connect the RP79xxA load unit.
Step 2. Turn on the unit under test and program the instrument settings as described in the test record
form under “Voltage Programming & Readback, Min Voltage”. Turn the output on. The output status
should be “CV”, with the output current close to zero.
Step 3. Record the output voltage reading from the DMM (Vout) and the voltage measured by the
instrument over the interface. The readings should be within the limits specified in the test record form
for the appropriate model under “Voltage Programming & Readback, Minimum Voltage”.
The next step applies the full output voltage of the unit under test to the output.
Step 4. Program the instrument settings as described in the test record form under “Voltage
Programming & Readback, High Voltage”.
Step 5. Record the output voltage reading from the DMM (Vout) and the voltage measured by the
instrument over the interface. The readings should be within the limits specified in the test record form
for the appropriate model under “Voltage Programming & Readback, High Voltage”.
Constant Voltage Load Effect
This test measures the change in voltage resulting from a change in current from full load to no load.
Step 1. Turn off the unit under test. Connect the DMM to the sense screws and connect the RP79xxA
load unit (see
).
Step 2. Turn on the unit under test and program the instrument settings as described in the test record
form under “CV Load Effect”. Turn the output on.
Step 3. Set the RP79xxA load for the rated output current of the unit under test as described in the test
record form under “CV Load Effect”. Turn the output on. The output status of the unit under test should
be “CV”. If it isn’t, adjust the load so that the output current drops slightly.
Keysight N69100 Series Operating and Service Guide
231
6 Verification and Calibration
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......