FETCh[:SCALar]:CURRent:LOW?
FETCh[:SCALar]:VOLTage:LOW?
Returns the Low level of a pulse waveform. Values returned are either in amperes, or volts. See
.
Parameter
Typical Return
(none)
<LOW value>
Returns the measured low level voltage FETC:VOLT:LOW?
FETCh[:SCALar]:CURRent:MAXimum?
FETCh[:SCALar]:CURRent:MINimum?
FETCh[:SCALar]:POWer:MAXimum]?
FETCh[:SCALar]:POWer:MINimum]?
FETCh[:SCALar]:VOLTage:MAXimum?
FETCh[:SCALar]:VOLTage:MINimum?
Returns the maximum or minimum value. Values returned are either in amperes, volts, or watts.
Parameter
Typical Return
(none)
<MIN value>
<MAX value>
Returns the measured maximum current FETC:CURR:MAX?
Returns the measured minimum voltage FETC:VOLT:MIN?
Returns the measured maximum power FETC:POW:MAX?
FETCh:AHOur? [IGNORE_OVLD]
FETCh:WHOur? [IGNORE_OVLD]
FETCh:AHOur? - Returns the accumulated amp-hours.
FETCh:WHOur? - Returns the accumulated watt-hours.
See
Amp-Hour and Watt-Hour Measurements
for details.
If any measurement sample was overrange, the query returns SCPI Not a Number (9.91E37). If the
optional IGNORE_OVLD parameter is sent, the accumulated measurement will be returned even if
some samples were outside of the measurement range.
Parameter
Typical Return
IGNORE_OVLD ignore overload measurements
<amp-hours>
<watt-hours>
Returns the amp-hour measurement FETC:AHO?
Returns the watt-hour measurement FETC:WHO?
5 SCPI Programming Reference
166
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......