INSTrument:GROup:PRIMary:CONNect:MODE AUTO|MANual
INSTrument:GROup:PRIMary:CONNect:MODE?
Specifies the connection mode of the primary unit. This setting is saved in non-volatile memory.
AUTO - the primary unit will try to connect to the previously discovered secondary units at power-on.
MANual - the primary will connect to the previously discovered secondary units when it receives a
connection command from the front panel or from INST:GROUP:PRIM:CONN.
Parameter
Typical Return
AUTO | MANual
AUTO or MAN
Specifies automatic connection mode INST:GRO:PRIM:CONN:MODE AUTO
INSTrument:GROup:PRIMary:DISCover
Instructs the primary unit to discover all secondary units connected to the primary/secondary bus.
Parameter
Typical Return
(none)
(none)
Discover all secondary units INST:GRO:PRIM:DISC
INSTrument:GROup:PRIMary:RESet
Resets the primary-discovered secondary configuration. It disconnects any secondary-connected
units and returns them to independent operation.
Parameter
Typical Return
(none)
(none)
Resets all secondary units INST:GRO:PRIM:RES
INSTrument:GROup:SECondary:ADDRess <
value
>
Sets the secondary unit's bus address. Each secondary unit in a primary/secondary group must have a
unique bus address or bus communication will fail. This setting is saved in non-volatile memory.
Parameter
Typical Return
1 - 9
1
Set the secondary unit's address to 1 INST:GRO:SEC:ADDR 1
Keysight N69100 Series Operating and Service Guide
177
5 SCPI Programming Reference
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......