If the instrument fails any of the tests or if abnormal test results are obtained, try calibrating the unit. If
calibration is unsuccessful, return the unit to a Keysight Technologies Service Center.
Refer to the
Recommended Test Equipment and Setups
section for a list of the equipment and test
setups required for verification. Also refer to the
section for information about
connecting the voltmeter, current shunt, and load.
SHOCK HAZARD, LETHAL VOLTAGES - Some models generate output voltages up
to 2,000 VDC! Ensure that all instrument connections, load wiring, load and sense
connections are insulated using high-voltage wire. Always attach the safety cover
provided, so that no accidental contact with lethal voltages can occur.
Always turn off the output when connecting or disconnecting any equipment on the
sense or output terminals of the unit.
Equipment Damage It is recommended to set the over-voltage protection function
of the instrument slightly higher than its operating point during the verification
procedure. This will prevent damage to any external equipment (electronic load,
differential amplifier) that may occur if the output voltage is accidentally
programmed higher than the prescribed voltage setting.
Turn the unit off or send a Reset command after completing the verification
procedure to return all instrument settings to their default values.
Verification Setups
Models N69143A, N69145A, and N69146A
Keysight N69100 Series Operating and Service Guide
229
6 Verification and Calibration
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......