Test Equipment and Setups
Test Equipment
The test equipment recommended for the performance verification and adjustment procedures is
listed below. If the exact instrument is not available, substitute calibration standards of equivalent
accuracy.
Instrument
Requirements
Recommended Model
Use1
Digital Multimeter
Resolution: 10 nV @ 1V;
Readout: 8 1/2 digits
Accuracy: 20 ppm
Keysight 3458A
V, C
Current Shunt2
50 A (0.05
Ω
) 0.01%, TC=4ppm/
Ω
C
100 A (0.01
Ω
) 0.01%, TC=4ppm/
Ω
C
300 A (0.001
Ω
) 0.01%, TC=4ppm/
Ω
C
1000A, 0.1m
Ω
0.02%, TC=25ppm/
Ω
C
Guildline 9230A/50
Guildline 9230A/100
Guildline 9230A/300
Guildline 9230A/1000 or equivalent
V, C
Electronic load/source 2
20 V, 800 A, 10 kW
80 V, 250 A, 10 kW
160 V, 125 A, 10 kW
1000 V, 60 A, 20 kW
2000 V, 30 A, 20 kW
Keysight RP7943A
Keysight RP7945A
Keysight RP7946A
Keysight RP7972A
Keysight RP7973A
V
GPIB controller
Full GPIB capabilities
Keysight 82350B or equivalent
V, C
Oscilloscope
Sensitivity: 1 mV
Bandwidth: 20 MHz
Probe: 10:1 with RF tip
Probe: 100:1 with RF tip
Keysight DSO6054A or equivalent
Keysight 10073D
Keysight 10076C
V
RMS Voltmeter
True RMS
Bandwidth: 20 MHz
Sensitivity: 100 μV
Rhode and Schwartz URE3, Keysight
3458A, or equivalent
V
Voltage divider
5kV:5V (1000:1) Accuracy <0.01%
Ohms-Lab KVVB-5-5 or equivalent
V, C
Differential Amplifier
Bandwidth: 20 MHz
LeCroy 1855A, DA1850A, or equivalent
V
Differential Probe
10:1/100:1 selectable
LeCroy DXC100A, or equivalent
V
Terminations
50
Ω
BNC termination
V
1
V=Verification; C=Calibration
2
Dependent on rating of model being tested
6 Verification and Calibration
226
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......