OCF- Negative over-current fault is active in current priority mode, when the DUT applies a voltage to
the output terminals that is greater than the positive current limit setting. It minimizes the resulting
peak current at the output terminals by comparing the actual output current to a fixed level of 105% of
the unit's rated current. If the current exceeds this level, the unit will go into a latched protection.
Set the Over-Voltage Protection
The over-voltage protection will turn the output off if the output voltage reaches the programmed
over-voltage limit. The OVP circuit monitors the voltage at the + and – sense terminals. An OVP
shutdown will automatically occur if the + and – sense leads are accidentally shorted to each other.
Front Panel Menu Reference
SCPI Command
Select
Protect\OVP
Enter a value in the OVP level box. Then press
Select
.
To set the OVP level to 400 volts:
Set the Over-Current Protection
Enable OCP
With over-current protection enabled, the output turns off when the output current reaches the
current limit setting and transitions from constant voltage (CV) to current limit (CL+ or CL-).
Front Panel Menu Reference
SCPI Command
Select
Protect\OCP
Check
Enable OCP
. Then press
Select
.
To enable OCP:
Delay OCP
You can specify an OCP delay to prevent momentary output settings, load, and status changes from
tripping the over-current protection. In most cases these momentary conditions would not be
considered an over-current protection fault, and having an OCP condition disable the output when
they occur would be a nuisance. Specifying an OCP delay lets the OCP circuit ignore these momentary
changes during the specified delay period. Once the OCP delay time has expired and the over-current
condition persists, the output will shut down. The following selections control the start of the over-
current delay timer:
Settings change - this starts the over-current delay whenever a command changes the output
settings. It includes changes made by the transient system so that the timer is started at each list step
and at each Arb output change. It also includes voltage and current slew changes, so that the timer is
restarted throughout the entire slew time.
Current limit - starts the over-current delay timer by any transition of the output into current limit.
The delay can be programmed from 0 to 0.255 seconds. You can specify if the OCP delay timer is
started by any transition of the output into current limit, or only at the end of a settings change in
voltage, current, or output state.
4 Using the DC Power System
110
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......