Parameter
Typical Return
0.1% to 102% of rating, *RST 1% of rating
<voltage limit>
Sets the voltage limit to 20 V: VOLT:LIM 20
[SOURce:]VOLTage:MODE FIXed|STEP|LIST|ARB
[SOURce:]VOLTage:MODE?
Sets the transient mode. This determines what happens to the output voltage when the transient
system is initiated and triggered.
FIXed keeps the output voltage at its immediate value.
STEP steps the output to the triggered level when a trigger occurs.
LIST causes the output to follow the list values when a trigger occurs.
ARB causes the output to follow the arbitrary waveform values when a trigger occurs.
Parameter
Typical Return
FIXed|STEP|LIST|ARB, *RST FIXed
FIX, STEP, LIST, or ARB
Sets the voltage mode to Step: VOLT:MODE STEP
[SOURce:]VOLTage:PROTection[:LEVel] <
value
>|MIN|MAX
[SOURce:]VOLTage:PROTection[:LEVel]? [MIN|MAX]
Sets the over-voltage protection level. Units are in volts. If the output voltage exceeds the OVP level,
the output is disabled and the Questionable Condition status register OV bit is set.
Parameter
Typical Return
0 to 120% of rating, *RST 120% of rating
<protect level>
Sets the over-voltage protection to 24 V: VOLT:PROT 24
l
An over-voltage condition can be cleared with the OUTput:PROTection:CLEar command after the
cause of the condition has been removed.
[SOURce:]VOLTage:RESistance[:LEVel][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage:RESistance[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
Sets the output resistance level. Only applies in voltage priority mode. Units are in ohms. Refer to the
tables for the model-specific resistance programming ranges.
Parameter
Typical Return
0 to maximum (model dependent) | MIN | MAX, *RST 0
0
Specifies an output resistance of 0.5 ohms: VOLT:RES 0.5
l
When units are connected in parallel, the maximum output resistance is reduced. The pro-
grammable resistance for a single unit must be divided by the total number of paralleled units.
5 SCPI Programming Reference
204
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......