![Keysight Technologies N69100 Series Скачать руководство пользователя страница 190](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218190.webp)
SENSe Subsystem
Sense commands control the current measurement ranges and window as well as the data
acquisition sequence.
SENSe:AHOur:RESet
SENSe:WHOur:RESet
Resets the amp-hour or watt-hour measurement to zero.
Parameter
Typical Return
(none)
(none)
Resets the amp-hour measurement: SENS:AHO:RES
Resets the watt-hour measurement: SENS:WHO:RES
SENSe:FUNCtion:CURRent 0|OFF|1|ON
SENSe:FUNCtion:CURRent?
SENSe:FUNCtion:VOLTage 0|OFF|1|ON
SENSe:FUNCtion:VOLTage?
Enables or disables current or voltage measurements.
Parameter
Typical Return
0|OFF|1|ON, *RST ON
0 or 1
Enables current measurements: SENS:FUNC:CURR ON
Disables voltage measurements: SENS:FUNC:VOLT OFF
SENSe:SWEep:NPLCycles <
value
>|MIN|MAX
SENSe:SWEep:NPLCycles? [MIN|MAX]
Sets the measurement time in number of power line cycles. Increasing the number of power line
cycles reduces the measurement noise on current and voltage measurements. Changing the NPLC
changes the number of points and time interval setting.
Parameter
Typical Return
0.0003072 to 1,258,290,000,000, *RST 1
<NPLC value>
Specifies 100 power line cycles: SENS:SWE:NPLC 100
5 SCPI Programming Reference
190
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......