Test Record Forms - APS family
Keysight N69143A
N69143A
Test Record
Report Number ______
Date__________
Test Description
Model
Min. Specs
Results
Max. Specs
Voltage Programming & Readback
Minimum voltage (Vout):
Voltage measured over interface:
High voltage (Vout):
Voltage measured over interface:
Both
Both
Both
Both
0.015996
Vout - 0.004004
19.992
Vout - 0.008
_______
_______
_______
_______
0.024004
Vout + 0.004004
20.008
Vout + 0.008
CV Load Effect:
Both
- 0.001
_______
+ 0.001
CV Ripple and Noise
peak-to-peak:
rms:
Both
Both
N/A
N/A
_______
_______
0.030
0.003
Transient response @ 900 µs:
Both
- 0.2
_______
+ 0.2
Current Programming & Readback
Minimum current (Iout):
Current measured over interface:
High current (Iout):
Current measured over interface:
Both
Both
Both
Both
- 0.37
Iout - 0.37
798.91
Iout - 1.09
_______
_______
_______
_______
0.37
Iout + 0.37
801.09
Iout + 1.09
CC Load Effect:
Both
- 0.050
_______
+ 0.050
Current Sink Tests
10% of current rating:
Both
- 82.0434
_______
- 81.1566
N69143A UUT Settings
RP7943A Load Settings
Voltage Programming & Readback Min:
Voltage Programming & Readback High:
CV Load Effect, CV Ripple and Noise:
Transient Response:
Current Programming & Readback Min:
Current Programming & Readback High:
CC Load Effect:
10% of current rating (Isink):
Voltage priority: 0.02 V, 408 A
Voltage priority: 20 V, 408 A
Voltage priority: 10 V, 816 A
Voltage priority: 10 V, 816 A
Current priority: 10.2 V, 0 A
Current priority: 10.2 V, 800 A
Current priority: 20.4 V, 400 A
Voltage priority: 0 V, 0 A
not used
not used
Current priority: 10.2 V, -800 A
Current priority: 15 V, -320 A to -720 A, slew=10 A/µs
Voltage priority: 10 V, 0 A
Voltage priority: 10 V, -816 A
Voltage priority: 20 V, -408 A
Voltage priority: 10 V, 800 A
Keysight N69100 Series Operating and Service Guide
239
6 Verification and Calibration
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......