FUNCtion Command
[SOURce:]FUNCtion CURRent|VOLTage
[SOURce:]FUNCtion?
Sets the output regulation - voltage priority or current priority. In voltage priority mode, the output is
controlled by a constant voltage feedback loop, which maintains the output voltage at its
programmed setting. In current priority mode, the output is controlled by a constant current feedback
loop, which maintains the output current at its positive or negative programmed setting.
Refer to
for more information.
Parameter
Typical Return
CURRent|VOLTage, *RST VOLTage
CURR or VOLT
Sets the output regulation to current priority: FUNC CURR
5 SCPI Programming Reference
168
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......