![Keysight Technologies N69100 Series Скачать руководство пользователя страница 180](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218180.webp)
[SOURce:]LIST:TOUTput:BOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON}
[SOURce:]LIST:TOUTput:BOSTep[:DATA]?
[SOURce:]LIST:TOUTput:EOSTep[:DATA] 0|OFF|1|ON {,0|OFF|1|ON}
[SOURce:]LIST:TOUTput:EOSTep[:DATA]?
Specifies which list steps generate a trigger signal at the beginning of step (BOSTep) or end of step
(EOSTep). A trigger is only generated when the state is set to ON. The trigger signal can be used as a
trigger source for measurements and transients of other units, and for digital port pins configured as
trigger outputs.
Parameter
Typical Return
0|OFF|1|ON
0 or 1
To generate triggers at the beginning of the second step of a 3-step list:
LIST:TOUT:BOST OFF,ON,OFF
5 SCPI Programming Reference
180
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......