Set the Output Turn-On/Turn-Off Mode
The turn-on/turn-off setting only applies when the unit is operating in voltage priority
mode. In current priority mode, the turn-on/turn-off behavior is set to High impedance.
The voltage priority turn-on and turn-off behavior can be set to either high or low impedance mode.
Low impedance mode is for devices like power converters. During output transitions, current is
sourced or discharged for fast output voltage response.
High impedance mode is for devices like batteries, where output transitions are controlled to minimize
output current.
Front Panel Menu Reference
SCPI Command
Select
Output\Advanced\Tmode
.
Select either High impedance or Low impedance for
turn-on an turn-off modes.
Check Coupled to couple the turn-on and turn-off
modes. Then press
Select
.
To select high impedance turn-on:
To select low impedance turn-off:
To couple the turn-on and turn-off modes:
Enable the Output
SHOCK HAZARD, LETHAL VOLTAGES - Some models generate output voltages up to
2,000 VDC! Ensure that all instrument connections, load wiring, load and sense
connections are insulated using high-voltage wire. Always attach the safety cover
provided, so that no accidental contact with lethal voltages can occur.
Because of internal circuit start-up procedures and any installed relay options, OUTPut ON may take
tens of milliseconds to complete its function. OUTPut OFF delays may also be in effect. For more
information regarding output turn-on and turn-off delays, refer to
.
Front Panel Menu Reference
SCPI Command
Press the
On/Off
key.
In addition to the front panel and SCPI Output On and Output Off commands, you can also use
OnCouple and OffCouple signals to enable and disable the output. Refer to
for
more information. When coupling is enabled, changing the turn-on setting also changes the turn-off
setting and vice-versa.
During a 1-cycle AC line dropout the unit may reboot. The output will remain off
after reboot until the operator reinstates the previous settings, either by the front
panel controls or using a computer program. This behavior is consistent with safe
operating procedures.
4 Using the DC Power System
100
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......