OUTPut Subsystem
The Output subsystem controls the output state, power-on, protection, and relay functions.
OUTPut [:STATe] 0|OFF|1|ON
OUTPut[STATe]?
SHOCK HAZARD, LETHAL VOLTAGES - Some models generate output voltages up to
2,000 VDC! Ensure that all instrument connections, load wiring, load and sense
connections are insulated using high-voltage wire. Always attach the safety cover
provided, so that no accidental contact with lethal voltages can occur.
Enables or disables the output. The state of a disabled output is a condition of zero output voltage
and zero source current.
Parameter
Typical Return
0|OFF|1|ON, *RST OFF
0 or 1
Turns the output off: OUTP OFF
l
When output is enabled, the front-panel status indicator changes from OFF to indicate the oper-
ating status of the instrument (CV, CC, etc.).
l
Separate delays can be programmed for the off-to-on and the on-to-off transition using
OUTPut:DELay:RISE and OUTput:DELay:FALL.
l
Because of internal circuit start-up procedures, OUTPut ON may take 8 milliseconds to complete
its function in voltage priority mode, and 7 milliseconds in current priority mode.
OUTPut[:STATe]:COUPle[:STATe] 0|OFF|1|ON
OUTPut[:STATe]:COUPle[:STATe]?
Enables or disables output coupling. Output coupling allows the outputs of multiple instruments to
turn on and off sequentially according to their specified OUTPut:DELay:RISE and OUTput:DELay:FALL
programming delays. This parameter is saved in non-volatile memory.
Parameter
Typical Return
0|OFF|1|ON
0 or 1
Turns the output coupling state on: OUTP:COUP ON
l
You must connect and configure the ONCouple and OFFCouple digital connector pins of all syn-
chronized instruments as described in the
section.
l
Because some power supplies have different minimum delay offsets, you must also specify a com-
mon delay offset for all of the synchronized units. This value must be the largest delay offset of the
synchronized group. Use OUTPut:COUPle:MAX:DOFFset? to query the delay offset for each unit.
The largest value returned must be specified as the common delay offset for each unit.
5 SCPI Programming Reference
184
Keysight N69100 Series Operating and Service Guide
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......