INITiate Subsystem
Initiate commands arm the trigger system. This moves the trigger system from the "idle" state to the
"wait-for- trigger" state; which enables the instrument to receive triggers. An event on the selected
trigger source causes the trigger to occur.
INITiate[:IMMediate]:ACQuire
INITiate[:IMMediate]:TRANsient
INITiate:ACQuire - Initiates the measurement trigger system.
INITiate:TRANsient - Initiates the transient trigger system.
Parameter
Typical Return
(none)
(none)
Initiate the measurement-trigger system: INIT:ACQ
l
It takes a few milliseconds for the instrument to be ready to receive a trigger signal after receiving
the INITiate command.
l
If a trigger occurs before the trigger system is ready for it, the trigger will be ignored. Check the
WTG_meas bit in the operation status register to know when the instrument is ready.
l
Use ABORt commands to return the instrument to Idle.
INITiate:CONTinuous:TRANsient 0|OFF|1|ON
INITiate:CONTinuous:TRANsient?
Continuously initiates the transient trigger system. This allows multiple triggers to generate multiple
output transients.
Parameter
Typical Return
0|OFF|1|ON, *RST ON
0 or 1
Continually initiates the output trigger system: INIT:CONT:TRAN ON
l
With initiate continuous disabled, the output trigger system must be initiated for each trigger using
the INITiate:TRANsient command.
l
ABORt:TRANsient does not turn off continuous triggers if INITiate:CONTinuous:TRANsient ON has
been programmed. In this case, first turn off continuous triggers before sending the ABORt com-
mand.
Keysight N69100 Series Operating and Service Guide
175
5 SCPI Programming Reference
Содержание N69100 Series
Страница 1: ...Keysight Advanced Power System DC Power Supply N69100 Series Operating and Service Guide...
Страница 2: ......
Страница 14: ......
Страница 50: ...Outline Diagram for 10 kW Models 1 Quick Reference 50 Keysight N69100 Series Operating and Service Guide...
Страница 51: ...Outline Diagram for 20 kW Models Keysight N69100 Series Operating and Service Guide 51 1 Quick Reference...
Страница 52: ......
Страница 80: ......
Страница 94: ......
Страница 210: ...Status Diagram 5 SCPI Programming Reference 210 Keysight N69100 Series Operating and Service Guide...
Страница 212: ...Trigger Diagram 5 SCPI Programming Reference 212 Keysight N69100 Series Operating and Service Guide...
Страница 230: ...Models N69172A and N69173A 6 Verification and Calibration 230 Keysight N69100 Series Operating and Service Guide...
Страница 244: ......