2-6
Preparing the Mass Spectrometer for Operation
To install the low-volume vials
:
1. If a standard reservoir bottle is fitted, remove it.
2. Screw the low-volume adaptors into the manifold and finger-tighten
them.
3. Screw the low-volume vials into the adaptors.
4. For each low-volume vial, ensure that the ends of the solvent delivery
tubes are positioned so that they are close to, but do not touch, the
bottom of the vial (see
Warning:
The reservoir bottles can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
Low-volume glass vials are fragile and can shatter,
cutting fingers. Take care when screwing them in, and never use
force.
Low-volume vial
Low-volume adaptor
Solvent delivery tube