4-10
Configuring the NanoLockSpray Source
2. Thread the indicated length of PTFE tubing through each of the
“finger-tight” nuts, and tighten them into the tee piece in the
arrangement as shown.
3. Using the SealTight nut and ferrule, connect the 0.023-inch ID PTFE
tubing to the pressure controller connection on the base of the source.
4. Using the nut and ferrule at the sprayer assembly, connect the
0.006-inch ID PTFE tubing to the sprayer.
Installing the glass capillary sprayer
Required materials
Chemical-resistant, powder-free gloves
To install the glass capillary sprayer:
1. Prepare the instrument for working on the source (see
2. Ensure the instrument is in Source Standby mode and that the status
indicator is yellow.
f
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The probe and source can be hot. To avoid burn injuries,
take great care while working with these components.
0.023-inch ID, 120-mm tubing
Pressure
controller
Sprayer
0.006-inch ID, 500-mm tubing
0.006-inch ID, 30-mm tubing