3-8
Configuring the LockSpray Source
To install the capillary:
1. Remove the existing capillary (see
).
2. Use the sharp knife or PEEK tubing cutter to cut an approximately
60-cm (24-inches) length of red, PEEK tubing.
Requirement:
Cut the tubing squarely (that is, perpendicular to its
horizontal axis).
3. Insert one end of the red, PEEK tubing in the probe inlet connector, and
screw the connector, finger-tight, into the PEEK “finger tight” nut and
ferrule.
Rationale:
Doing so ensures a minimum dead volume when fitting the
capillary.
Warning:
The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The ESI probe tip is sharp. To avoid puncture wounds, handle
the probe with care.
TP02671
PEEK “finger tight” nut and ferrule
PEEK tubing
PEEK union