Replacing the ion block source heater
5-95
3. Use the combined 2.5-mm Allen wrench and cone extraction tool to
loosen the 2 captive screws securing the ion block cover plate.
4. Remove the ion block cover plate.
5. Use the combined 2.5-mm Allen wrench and cone extraction tool to
remove the 2 screws securing the heater wires to the PEEK terminal
block.
Ion block cover plate securing screw
Ion block cover plate
Heater cartridge wire
securing screws
PEEK terminal block