5-92
Maintenance Procedures
2. Gripping the probe heater as shown, carefully pull it off the probe
assembly.
3. Dispose of the probe heater in accordance with local environmental
regulations.
Caution:
To avoid damaging the probe heater’s electrical
connections, do not twist the heater when removing it from the
probe assembly.
Warning:
The probe heater can be contaminated with
biohazardous and/or toxic materials. Dispose of it according
to local environmental regulations.