5-48
Maintenance Procedures
7. Grasp the isolation valve and pull it out of the ion block.
8. Use the O-ring removal kit to carefully remove the isolation valve O-ring
9. If the isolation valve O-ring shows signs of deterioration or damage,
dispose of it in accordance with local environmental regulations.
10. Use the combined 2.5-mm Allen wrench and cone extraction tool to
loosen the captive PEEK terminal block securing screw.
Warning:
The isolation valve O-ring can be contaminated
with biohazardous and/or toxic materials. Dispose of it
according to local environmental regulations.
Isolation valve
O-ring