Replacing the IntelliStart Fluidics tubing 5-115
10. Connect the grounded union to the source’s reference probe.
•
For the LockSpray source, use 1/16-inch PEEK tubing connected as
follows:
–
At the grounded union end, use a long “finger-tight” fitting.
Long “finger-tight” fitting:
–
At the source reference-probe connection, use the PEEK,
“finger-tight” nut and ferrule.
PEEK, “finger-tight” nut and ferrule:
•
For the NanoLockSpray source, see
11. Close the access doors to the IntelliStart Fluidics system.
Warning:
To avoid electric shock, do not use stainless steel tubing
to connect the grounded union to the reference probe.
Caution:
Ensure that the tubing does not become trapped when
you close the access door to the IntelliStart Fluidics system.