Maintenance Procedures
Installing the new TaperTip and capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
Fused-silica cutting tool
•
Pin-plug capillary locator
•
TaperTip
•
A 375-mm length of 25-µL, fused-silica capillary, sleeved for protection
by 0.015-inch ID PEEK tubing
To install the new TaperTip and capillary:
1. On the reference probe assembly, screw the
pin-plug
capillary locator
into the end of the union.
2. Thread the capillary through the body of the reference probe.
3. Slide a PEEK coupler over the end of the fused-silica capillary.
Pin-plug capillary locator
Reference-probe body
Union