Replacing the mass spectrometer’s air filters 5-105
Required materials
•
M4 POZIDRIV
®
screwdriver
•
New filters
To replace the air filters on the sides of the instrument:
For each filter:
1. Use the M4 POZIDRIV
screwdriver to remove the screws securing the
louvered panel to the instrument enclosure.
2. Remove the louvered panel from the instrument enclosure.
3. Remove the filter from the instrument.
4. Dispose of the filter.
5. Fit the new filter into the instrument.
6. Fit the louvered panel to the instrument enclosure.
7. Use the M4 POZIDRIV
screwdriver to fit and tighten the screws
securing the louvered panel to the instrument enclosure.