5-66
Maintenance Procedures
To fit the ESI probe tip and gasket:
1. Fit the new metal gasket into the probe tip.
2. Fit the probe tip over the capillary, and screw the tip onto the probe
assembly.
3. Use the 10-mm wrench to tighten the probe tip.
4. Use the nebulizer adjuster knob to adjust the capillary so that it
protrudes by approximately 0.5 mm from the end of the probe.
5. Fit the ESI probe to the source (see
).
Replacing the ESI probe sample capillary
Replace the stainless steel sample capillary in the ESI probe if it becomes
irreversibly blocked or if it becomes contaminated or damaged.
Removing the existing capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
7-mm wrench
•
8-mm wrench
•
10-mm wrench
•
Needle-nose pliers
Warning:
The probe and source components can be contaminated
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The ESI probe tip is sharp. To avoid puncture wounds, handle
the probe with care.
Caution:
To avoid gas leakage, fully tighten the probe tip.