Connecting the NanoLockSpray source camera
B-27
To connect the camera to the workstation:
1. Connect the coaxial cable between the video output connector on the
instrument’s rear panel and the interface box.
2. Connect the USB cable between the interface box and any free USB
socket on the workstation.
Installing the camera driver software
The workstation is supplied with the camera driver software pre-installed.
However, if you reinstall or reconfigure the workstation, you may need to
reinstall the camera driver software. If this is the case, then Windows
presents the Found New Hardware Wizard when you connect the USB
Converter.
Installing the camera driver software:
1. In the Found New Hardware Wizard dialog, prevent Windows from
connecting to Windows Update by selecting No, not at this time.
2. Click Next.
3. Click Install from a list or specific location (Advanced).
4. Click Next.
5. Select Search for the best driver in these locations.
6. Select Include this location in the search and:
•
For MassLynx™, browse to C:\MassLynx\USBCameraDriver.
•
For UNIFI™, browse to
C:\Program Files\Waters\UNIFI\USBCameraDriver.
7. Click Next.
Result:
The software installs.
8. Click Finish.