5-86
Maintenance Procedures
To remove the reference probe:
1. Prepare the instrument for working on the source (see
2. Disconnect the reference probe’s cable from the instrument’s high
voltage connector.
3. Disconnect the fused silica capillary from the instrument’s IntelliStart
Fluidics system.
4. Use the combined 2.5-mm Allen wrench and cone extraction tool to
unscrew the 3 reference probe retaining screws.
5. Remove the reference probe from the probe adjuster assembly.
Warning:
All plumbing connections, the probe, and the source can
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this procedure.
Warning:
The reference probe tip consists of 5 cm of exposed,
fused-silica TaperTip, which is sharp and fragile. To avoid
puncture wounds, handle the probe with care.
Combined 2.5-mm Allen wrench and cone extraction tool
Reference probe retaining screw