5-12
Maintenance Procedures
Corona pin mounting contact:
4. Fit the corona pin to the corona pin mounting contact, ensuring that the
corona pin is securely mounted and that its tip aligns with the sample
cone orifice.
Corona pin:
Warning:
The corona pin tip is sharp. To avoid puncture wounds,
handle it with care.
TP03130
Corona pin mounting
contact blanking plug
TP03130
Corona pin
Sample cone orifice