B-22 External Connections
External connections 1:
Pin
Function
Rating
1
Event In 1+, digital signal,
o3.3V max +5V
+5 V
2
Event In 1-, digital ground, 0V
0 V
3
Not used
4
Event In 2+, digital signal,
o3.3V max +5V
+5 V
5
Event In 2-, digital ground, 0V
0 V
6
Not used
7
Not used
8
CE Interlock Out, common
+30 V d.c., 100 mA
9
CE Interlock Out, normally closed (N/C)
+30 V d.c., 100 mA
10
CE Interlock Out, normally open (N/O)
+30 V d.c., 100 mA
External connections 2:
Pin
Function
Rating
1
Analog Out +, instrument-driven electrical
output
None
2
Analog Out -, ground
None
3
Gas Fail Interlock, common
+30 V d.c., 100 mA
4
Gas Fail Interlock, normally closed (N/C)
+30 V d.c., 100 mA
5
Gas Fail Interlock normally open (N/O)
+30 V d.c., 100 mA
6
Event Out 1 +
+30 V d.c., 100 mA
7
Event Out 1 -
+30 V d.c., 100 mA
8
Not used
9
Event Out 2 +
+30 V d.c., 100 mA
10
Event Out 2 -
+30 V d.c., 100 mA