3-2
Configuring the LockSpray Source
Configuring the LockSpray source
The following table summarizes how you configure the LockSpray source for
the various ionization modes.
Configuring for ESI mode
To operate in ESI mode, you must fit the ESI probe to the LockSpray source
enclosure. If you intend using the small-bore capillary option, fit the capillary
).
For more information on using ESI mode, see the Xevo G2 QTof system online
Help.
Installing the ESI probe
Required materials
•
Chemical-resistant, powder-free gloves
•
Sharp knife or PEEK™ tubing cutter
Configuring the LockSpray source:
Ionization mode
Probe type
Corona pin fitted?
ESI
ESI
No
APCI
APCI
Yes
ESCi
ESI
Yes