Replacing the IntelliStart Fluidics tubing 5-121
7. Using two long “finger-tight” fittings, connect the red, 1/16-inch,
200-mm, PEEK tubing from port 3 of the sample selector valve to port 3
of the diverter valve.
8. Using a long “finger-tight” fitting, connect 1/16-inch, 1000-mm,
natural-colored PEEK tubing to port 4 of the diverter valve, and thread
it securely into the waste port.
Tip:
This is the same waste port as used in
A
B
C
A
B
C
A
B
C
Waters
A
B
C
Diverter valve
A
B
C
A
B
C
A
B
C
Waters
A
B
C
Waste port