Replacing the IntelliStart Fluidics tubing 5-113
8. Using a long “finger-tight” PEEK nut and 1/32-inch PEEK ferrule,
connect the red, 1/32-inch, 200-mm, PEEK tubing from port 4 of the
lock-spray selector valve to the left-hand side of the flow sensor, where
you use the 1/32-inch, 6-40 Valco™ compression fitting assembly.
Long “finger-tight” PEEK nut and 1/32-inch, PEEK ferrule:
1/32-inch, 6-40 Valco compression fitting assembly:
Caution:
The narrower, 1/32-inch tubing can slip out of the port.
Once the fitting is tightened, give the tubing a gentle tug to
ensure that it is gripped correctly.
A
B
C
A
B
C
A
B
C
Waters
A
B
C
Flow sensor
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...