Preparing the IntelliStart Fluidics system
2-5
To install the reservoir bottles:
1. Remove the reservoir bottle caps.
2. Screw the reservoir bottles onto the mass spectrometer, as shown below.
3. For each reservoir bottle, ensure that the ends of the solvent delivery
tubes are positioned so that they are close to, but do not touch, the
bottom of the bottle (see
).
Warning:
The reservoir bottles can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Reservoir bottle
Solvent delivery tube