Preparing the IntelliStart Fluidics system
2-7
Adjusting the solvent delivery tube positions
For correct operation of the IntelliStart Fluidics system, you must adjust each
solvent delivery tube so that its end is close to, but does not, touch, the bottom
of the reservoir bottle or low volume vial.
To adjust the position of a solvent delivery tube:
1. Open the access door to the fluidics pump (see the figure on
).
2. Loosen the finger-tight fitting for the solvent delivery tube you are
adjusting.
3. Move the solvent delivery tube so that its end is close to, but does not
touch, the bottom of the reservoir bottle or low volume vial.
4. Tighten the finger-tight fitting.
5. Close the access door.
Finger-tight fitting
Solvent delivery tube