B-26 External Connections
Connecting to the electricity source
The mass spectrometer requires a separate, grounded electricity source. The
ground connection in the electrical outlet must be common and connected near
the system.
To connect to the electricity source:
Recommendation:
Use a line conditioner or an uninterruptible power supply
(UPS) for optimum long-term input voltage stability.
1. Connect the female end of the power cord to the receptacle on the rear
panel of the mass spectrometer (see the figure on
).
2. Connect the male end of the mass spectrometer power cord to a suitable
200 to 240 V a.c. wall outlet.
Connecting the NanoLockSpray source camera
The NanoLockSpray™ source camera is connected to the workstation PC
through a USB converter.
Warning:
To avoid electrical shock, use the SVT-type power cord in the
United States and HAR-type (or better) in Europe. For information
regarding what cord to use in other countries, contact your local Waters
distributor.
External Connections 2
10
1
External Connections 1
10
1
System Activity
1
2
3
4
5
6
7
8
9
10
+
-
Event In 1
External Connections 1
Not used
CE
Interlock
OUT
+
-
Event In 2
Not used
-
IN
IN
Com
N/C
N/O
Com
N/C
N/O
1
2
3
4
5
6
7
8
9
10
+
-
Analogue
Out
External Connections 2
OUT
Gas Fail
Interlock
Not used
-
OUT
OUT
+
-
Event Out 1
+
-
Event Out 2
OUT
-
Video Output
Service Bus
Instrument LAN
EPC Com Port
EPC Parallel Port
Video output connector
Workstation
USB converter
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...