B-18 External Connections
Connecting the EPC
In addition to a workstation PC or instrument server, the Xevo G2 Tof has an
external EPC with a real-time operating system that provides the real-time,
data-processing required by the analog-digital controller (ADC).
Mass spectrometer rear panel connections to the EPC:
Requirement:
Use shielded network cables with the mass spectrometer to
ensure compliance with FCC limits.
To connect the EPC:
1. Connect one end of the 25-way, EPC cable to the 25-way, D-type
connector on the instrument.
2. Connect the free end of the 25-way, EPC cable to the 25-way COM port
D-type connector on the EPC.
Video Output
Service Bus
Instrument LAN
EPC Com Port
External Connections 2
10
1
External Connections 1
10
1
System Activity
200 - 240V, 50/60Hz, 1.1kW
1
2
3
4
5
6
7
8
9
10
+
-
Event
In
1
External Connections 1
Not
used
CE
Interlock
OUT
+
-
Event
In
2
Not
used
-
IN
IN
Com
N/C
N/O
Com
N/C
N/O
1
2
3
4
5
6
7
8
9
10
+
-
Analogue
Out
External Connections 2
OUT
Gas
Fail
Interlock
Not
used
-
OUT
OUT
+
-
Event
Out
1
+
-
Event
Out
2
OUT
-
EPC Parallel Port
ADC Ethernet
Power
EPC parallel port
EPC parallel port
EPC
Instrument LAN
ADC Ethernet
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...