5-82
Maintenance Procedures
Replacing the reference probe capillary (LockSpray
source)
Replace the reference probe capillary if it becomes blocked, and you cannot
clear it, or if it becomes contaminated or damaged.
When replacing the reference probe capillary, you can also replace the
microfilter. The microfilter is provided, with manufacturer’s instructions for
fitting it, in the spares kit for the reference probe assembly.
Removing the existing capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
4-mm wrench
•
Spares kit for the reference probe assembly
•
Extender tool for the SealTight nut (included in the spares kit)
SealTight
nut
Reference sprayer
(behind baffle)
Microfilter
Reference
probe
capillary