Replacing the IntelliStart Fluidics tubing 5-119
3. Using a long “finger-tight” fitting, connect 1/16-inch, 1000-mm,
natural-colored PEEK tubing to port 1 of the sample selector valve, and
thread the tubing into the waste port.
Long “finger-tight” fitting:
Tip:
The liquid waste system collects waste without requiring a fitted
connection. Waste drains through the connector at the base of the
instrument and into the waste bottle (see
).
4. Using a long “finger-tight” fitting, connect orange, 1/16-inch, 1000-mm,
PEEK tubing from port 2 of the sample selector valve to the wash bottle.
Requirement:
Push the tubing to the bottom of the wash solution, and
secure it to prevent the tubing floating to the surface during use.
A
B
C
A
B
C
A
B
C
Waters
A
B
C
Waste port