Connecting the liquid waste line
B-15
6. Connect one end of the tubing to the exhaust port on the instrument’s
rear panel.
7. Connect the free end of the tubing to the inlet port on the nitrogen
exhaust trap bottle.
8. Cut a second length of 12-mm tubing long enough to connect the
nitrogen exhaust trap bottle to the exhaust vent.
9. Insert one end of the tubing into the outlet port on the nitrogen exhaust
trap bottle.
10. Route the free end of the tubing to the exhaust vent.
Connecting the liquid waste line
Required materials
•
Chemical-resistant, powder-free gloves
•
Waste container
To connect the liquid waste line:
1. Place a suitable waste container below the mass spectrometer.
2. Slide a drain line over the barbed fitting of the drain (located at the
bottom, left-hand side of the mass spectrometer).
Caution:
To avoid gas leaks, use the sharp knife to cut the PTFE
tubing squarely (that is, perpendicular to its horizontal axis).
Warning:
The waste line and connection can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Caution:
To avoid distorting the drip tray or causing the drain cup
to leak, restrain the drain cup when attaching or removing the
waste line.
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...