5-116 Maintenance Procedures
Plumbing the IntelliStart Fluidics sample delivery system
This section explains how to plumb the sample delivery system for standard
flow applications. For low flow, nanoACQUITY applications, the column outlet
passes directly to the NanoFlow sprayer, bypassing the IntelliStart Fluidics
system.
IntelliStart Fluidics sample delivery tubing schematic:
All tubing is PEEK, with the exception of the connection between the pump
and sample selector valve, which is stainless steel.
Tubing dimensions—sample delivery system:
Valve/port
Connection
ID
(inch)
OD
(inch)
Color
Length
(mm)
Sample/1
Waste reservoir
0.040
1/16
Natural
1000
Sample/2
Wash reservoir
0.020
1/16
Orange
1000
Sample/3
Diverter valve
0.005
1/16
Red
200
Sample/4
Reservoir Bottle A
0.020
1/16
Orange
680
Sample/5
Reservoir Bottle B
0.020
1/16
Orange
680
Sample/6
Reservoir Bottle C
0.020
1/16
Orange
680
Sample/7
Analyte pump
0.040
1/16
n/a
500
Divert/4 Waste
0.040
1/16
Natural
1000
A
B
C
2
3
4
1
2
1
6
5
4
3
7
From pump
From wash bottle
To probe
From UPLC
Waste
port
Tubing
guides
Sample selector valve
Diverter
valve
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...