Configuring for ESI mode
3-3
To install the ESI probe:
1. Prepare the instrument for working on the source (see
2. Remove the protective sleeve, if fitted, from the ESI probe tip.
3. With the probe label facing you, carefully slide the ESI probe into the
hole in the probe adjuster assembly, ensuring that the probe location
dowel aligns with the location hole in the probe adjuster assembly.
Warning:
The LC system connections, ESI probe, and source can
be contaminated with biohazardous and/or toxic materials. Always
wear chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
To avoid electric shock, ensure that the instrument is
prepared for working on the source before commencing this procedure.
Warning:
The ESI probe tip is sharp. To avoid puncture wounds,
handle the probe with care.
TP03129
Location hole of the probe adjuster assembly
Probe location dowel
Probe label